大孔径CT模拟定位机扫描条件对CT值影响分析
引用本文:张坤,李芳,陈力,冯冰,李涛,王学敏,吴湘阳. 大孔径CT模拟定位机扫描条件对CT值影响分析[J]. 现代肿瘤医学, 2021, 0(9): 1575-1579. DOI: DOI:10.3969/j.issn.1672-4992.2021.09.023
作者姓名:张坤  李芳  陈力  冯冰  李涛  王学敏  吴湘阳
作者单位:1.陕西省肿瘤医院放疗科,陕西 西安 710061;2.西安交通大学第二附属医院,陕西 西安 710004
基金项目:陕西省卫建委项目(编号:2018A007)。
摘    要:目的:分析大孔径CT模拟定位机扫描条件对CT值的影响,为模拟定位工作中不同扫描条件的选择提供参考.方法:采用飞利浦大孔径CT模拟定位机扫描CIRS-062M电子密度模体并建立CT值-相对电子密度曲线.分析管电流、管电压、层厚、层间距、分辨率、准直器、旋转时间、重建模型、过滤器、几何位置多种扫描条件对CT值的影响.结果:...

关 键 词:放射治疗  CT模拟定位机  扫描协议  相对电子密度

Analysis of the effect of scanning conditions of Brilliance CT Big Bore on CT value
ZHANG Kun,LI Fang,CHEN Li,FENG Bing,LI Tao,WANG Xuemin,WU Xiangyang. Analysis of the effect of scanning conditions of Brilliance CT Big Bore on CT value[J]. Journal of Modern Oncology, 2021, 0(9): 1575-1579. DOI: DOI:10.3969/j.issn.1672-4992.2021.09.023
Authors:ZHANG Kun  LI Fang  CHEN Li  FENG Bing  LI Tao  WANG Xuemin  WU Xiangyang
Affiliation:1.Department of Radiation Oncology,Shaanxi Province Cancer Hospital,Shaanxi Xi'an 710061,China;2.The Second Affiliated Hospital of Xi'an Jiaotong University,Shaanxi Xi'an 710004,China.
Abstract:Objective:To analyze the impact of the scanning conditions of Brilliance CT Big Bore on CT value,and provide a reference for the selection of different scanning conditions in the simulation positioning work.Methods:The Philips Brilliance CT Big Bore was used to scan the CIRS-062M electron density phantom and establish the CT value-electron density curve.The effects of tube current,tube voltage,thickness,increment,resolution,collimator,rotation time,reconstruction,filter,geometric position on the CT value were analyzed.Results:The tube voltage change had the most significant effect on the CT value.The average deviation of the CT value of the 90 kV and 140 kV tube voltage was 66.1 HU and-24.0 HU.The difference was mainly concentrated in the high-density area.The filters YA and YB had a significant effect on the CT value,and the average deviation of CT values was-40.7 HU and-39.9 HU.Different scanning positions had a certain influence on the CT value.Conclusion:The influence of tube voltage,filter and scanning position on the CT value cannot be ignored.In clinical work,the use of the above parameters must be regulated to avoid deviations in dose calculation.
Keywords:radiotherapy  CT simulation machine  scanning protocol  relative electron density