Abstract: | To measure regional cerebral blood flow (rCBF) by blood sampling using 99mTc-ECD we devised a method of measuring the radioactive concentration in arterial blood sample with a gamma camera. In this method the head and a blood sample are placed within the same visual field to record the SPECT data of both specimens simultaneously. The results of an evaluation of the counting rate performance, applying the 30 hours decaying method using 99mTc solution showed that this method is not comparable to the well-type scintillation counter and in clinical cases the active concentration in arterial blood sample remained well within the dynamic range. In addition, examination of the influence of scattered radiation from the brain by the dilution method showed that it was negligible at a distance of more than 7.5 cm between the brain and the arterial blood sample. In the present study we placed a head-shaped phantom next to the sample. The results of the examinations suggested that this method is suitable for clinical application, and because it does not require a well-type scintillation counter, it is expected to find wide application. |