The influence of exposure parameters on jawbone model accuracy using cone beam CT and multislice CT |
| |
Authors: | Vandenberghe B,Luchsinger S,Hostens J,Dhoore E,Jacobs R SEDENTEXCT Project Consortium |
| |
Affiliation: | Department of Prosthetic Dentistry, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Katholieke Universiteit Leuven, Belgium. bart.vandenberghe@med.kuleuven.be |
| |
Abstract: | ObjectiveThe main purpose of this study was to investigate the influence of exposure parameters on jawbone model accuracy when using cone beam CT (CBCT) and multislice CT (MSCT).MethodsA lower and an upper edentulous human cadaver jaw were scanned using micro-CT (Skyscan 1173 high energy spiral scan micro-CT; Skyscan NV, Kontich, Belgium) at 35 μm to serve as true reference. The in vitro samples were exposed using six CBCT units and one MSCT system. CBCT exposure protocols were chosen according to clinically available settings. The variables were kilovoltage, milliamperage, voxel size and/or scan time. Image segmentation was based on local thresholds using profile lines. The resulting jawbone segmentations were registered with the reference and image processing was done to internally fill the segmentations. A point-based distance calculation was performed between the three-dimensional objects and reference scans and deviation percentages were calculated for 2 mm, 1 mm and 0.5 mm intervals.ResultsAll points of the MSCT surface models lay within a 1 mm deviation range and 98.5% within 0.5 mm compared with micro-CT. For the different CBCT systems, accuracy came close to MSCT with mean percentages of 98.9% within 1 mm deviation and 92.8% within 0.5 mm. A difference of approximately 1% between lower and upper jaws could be perceived. For the specific CBCT exposure protocols, only scan time and voxel size revealed certain significant differences.ConclusionJawbone model accuracy using CBCT was comparable with MSCT. The surface models of the upper jaws deviated slightly more than those for lower jaws. CBCT exposure settings had a limited influence on accuracy with scan time and voxel size as the main factors. |
| |
Keywords: | cone beam computed tomography surface model accuracy segmentation exposure protocol |
本文献已被 PubMed 等数据库收录! |
|