首页 | 本学科首页   官方微博 | 高级检索  
     


Transcutaneous Bilirubinometry
Authors:YOSHITADA YAMAUCHI  ITSURO YAMANOUCHI
Affiliation:Department of Paediatrics, Children's Medical Center, Okayama National Hospital, Okayama, Japan
Abstract:ABSTRACT. A total of 576 transcutaneous bilirubin measurements were performed on 336 Japanese full-term breast-fed newborn infants during the first twelve days of life. Our present study revealed that transcutaneous bilirubin measurements obtained from the forehead, chest, and sternum correlated well with serum bilirubin concentrations measured by AO bilirubinometer (0.910–0.922, p <0.001, n =576). The 95% confidence lmits were ±3.04 mg/dl for the forehead, ±2.85 mg/dl for the chest, and ±2.84 mg/dl for the sternum readings. The overall mean of values from the forehead, chest and sternum, when compared with individual means, was found to correlate better with serum bilirubin concentrations ( r =0.930, p <0.001, n =576) and improve the 95% confidence limits to ±2.68 mg/dl. These results demonstrated that the accuracy and reliability of TcB measurement could be increased further with multiple site measurement. The study clearly indicates that transcutaneous bilirubinometry is useful for clinical screening of serum bilirubin levels in Japanese full-term newborn infants.
Keywords:newborn    transcutaneous bilinrbinornetry    TcB index    hyperbilirubinemia