Safety and Efficacy of Pacing for Ventricular Tachycardia |
| |
Authors: | L.K. HOLLEY,&dagger ,M. COOPER,J.B. UTHER,D.A. ROSS |
| |
Affiliation: | Telectronics Pty. Limited, Sydney Australia;Cardiology Department, Westmead Hospital, Sydney, Australia |
| |
Abstract: | This study was undertaken to determine the safety and efficacy of three different pacing modalities on the termination of ventricular tachyarrhythmias. Thirty-two patients were studied in the electrophysiology laboratory. Three randomized pacing modalities were selected for attempted conversion: auto increment, auto burst, and random burst. In all three groups, arrhythmias with cycle lengths shorter than 230 ms required DC shock, with one exception. Those longer than 230 ms were terminated by pacing in 85% of cases. There was a 15% rate of acceleration. Thus, antitachycardia pacing for ventricular tachyarrhythmias should be considered only with defibrillating back-up. |
| |
Keywords: | antitachycardia pacing defibrillation |
|
|