首页 | 本学科首页   官方微博 | 高级检索  
     


Accuracy of recording horizontal condylar inclination and Bennett angle with the Cadiax compact
Authors:Celar A G  Tamaki K
Affiliation:Department for Fixed and Removable Prosthodontics, University of Vienna, Vienna, Austria. ales.celar@univie.ac.at
Abstract:This study assessed the accuracy of an electronic hinge axis tracing device (Cadiax compact) in measuring the horizontal condylar inclination (HCI) and the Bennett angle. The hinge axis movements were simulated with an articulator to which the hinge axis tracer was attached. On the articulator the pre-set HCI values were 20 degrees, 40 degrees and 60 degrees, the pre-set Bennett angles were 0 degrees, 5 degrees, 10 degrees, 15 degrees, 20 degrees. The maximum measuring errors ranged from 0 to 3.4 degrees (in average 1.2 degrees ) and yielded statistically significant differences between articulator setting and Cadiax compact measurement (P < 0.05) except the HCI of 40 degrees and the Bennett angle of 0 degrees. Because of the small range of the maximum measuring error the Cadiax compact represented reasonable accuracy for clinical application in anterior guidance restorations.
Keywords:horizontal condylar inclination    Bennett angle    hinge axis
本文献已被 PubMed 等数据库收录!