Biomechanical comparison of the FasT-Fix meniscal repair suture system with vertical mattress sutures and meniscus arrows |
| |
Authors: | Borden Peter Nyland John Caborn David N M Pienkowski David |
| |
Affiliation: | Division of Sports Medicine, Department of Orthopaedic Surgery, University of Louisville, Louisville, Kentucky 40202, USA. |
| |
Abstract: | BACKGROUND: A meniscal repair technique that combines the strength of vertical mattress sutures and the decreased tissue morbidity of an all-inside technique would be advantageous. HYPOTHESIS: The FasT-Fix Meniscal Repair Suture System will provide load at failure, stiffness, and displacement equivalent to that of vertical mattress sutures and superior to that of Meniscus Arrows. STUDY DESIGN: In vitro biomechanical study. METHODS: After repair of a 2-cm vertical longitudinal medial meniscal lesion, three groups of six human cadaveric knees were biomechanically tested in a random order on a servohydraulic device, and three groups of five specimens underwent cyclic loading. RESULTS: Specimens repaired with Meniscus Arrows had reduced load at failure, stiffness, and displacement, but there were no differences between the FasT-Fix and vertical mattress suture methods. During cyclic loading, specimens repaired with two Meniscus Arrows failed before test completion, whereas specimens repaired with two vertical mattress sutures (6.0 +/- 3.7 mm) or with two FasT-Fix implants (5.1 +/- 1.4 mm) maintained fixation with comparable displacements. CONCLUSIONS: The FasT-Fix provided load at failure, stiffness, and displacement comparable with that of vertical mattress sutures. Clinical Relevance: The results suggest that the FasT-Fix may be preferable to Meniscus Arrows for meniscal repair with minimal associated tissue morbidity. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|