直接测定尿中铅的石墨炉原子吸收光谱法
引用本文:侯晓燕.直接测定尿中铅的石墨炉原子吸收光谱法[J].预防医学文献信息,2010(7):637-639.
作者姓名:侯晓燕
作者单位:安阳市疾病预防控制中心,河南安阳455000
摘    要:目的]改进尿中铅的石墨炉原子吸收光谱法。方法]尿样加基体改进剂后,通过优化仪器实验条件,直接进样测定尿中铅。结果]尿铅浓度在0~120μg/L范围内呈良好的线性关系,相关系数r=0.9996,最低检出限为0.016μg/L,相对标准偏差小于3.08%,回收率为97.6%~103.2%。结论]该方法简便、快速、灵敏度高、精密度好、结果准确,适用于直接测定尿中铅。

关 键 词:尿    石墨炉原子吸收

Direct Determination of Lead in Urine by Graphite Furnace Atomic Absorption Spectrometry
HOU Xiao-yan.Direct Determination of Lead in Urine by Graphite Furnace Atomic Absorption Spectrometry[J].Liferatue and Information On Preventine Medicine,2010(7):637-639.
Authors:HOU Xiao-yan
Institution:HOU Xiao-yan. (Anyang Center for Disease Control and Prevention ,Anyang , 455000, Hgnan , China)
Abstract:Objective]To improve determination of lead in urine by graphite furnace atomic absorption spectrometry. Methods]Urine baseplates modifier, the experimental conditions by optimizing the device directly into the urine samples of lead. Results]Urinary lead concentration in the 0-120μg/L within the linear relationship, the correlation coefficient r = 0. 999 6 ,the lowest detection limit of 0. 016μg/L,relative standard deviation of less than 3.08% ,recoveries of 97.6% and 103.2 %. Conclusion]The method is simple, rapid, high sensitive, precise, accurate and suitable for direct determination of lead in urine.
Keywords:Urine  Lead  Graphite furnace atomic absorption
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号