Variation in the isocentre of a Philips Linear Accelerator (SL–20) used for stereotactic radiosurgery/stereotactic radiotherapy |
| |
Authors: | H D Souza,RC Joshi,T Ganesh,V Subramani,G Vasu,R Kumar,PK Julka,GK Rath |
| |
Affiliation: | H D’Souza,RC Joshi,T Ganesh,V Subramani,G Vasu,R Kumar,PK Julka,GK Rath |
| |
Abstract: | Stereotactic irradiation, either in the form of stereotactic radiosurgery (SRS) or stereotactic radiotherapy (SRT) of brain lesions requires high precision and submillimetre accuracy in the isocentre, the main determinants being gantry and couch rotations. It is thus necessary to evaluate the isocentre variation due to gantry and couch rotations in the particular setup for SRS/SRT. This paper describes variation in the isocentre of a Philips (now Elekta) SL-20 linear accelerator modified for adapting a couch-mounted radiosurgery system. By considering the isocentre as defined by a mechanical index as the standard, the variations in the isocentre of the linear accelerator were independently measured for the gantry and for couch rotations. The variation in the isocentre for gantry rotation was found to be between 0.1 mm and 0.9 mm, conforming to the submillimetre accuracy required for SRS/SRT. However, the isocentre variation due to couch rotation varied considerably, possibly because the couch is of the RAM type. The isocentre variation due to couch rotation is rectified by microadjusting the couch mount at the time of treatment using a laser target localizing frame. It is our conclusion that a modified linear accelerator can be used for performing SRS/SRT after careful and separate evaluation of the isocentre stability due to gantry and couch rotations. |
| |
Keywords: | isocentre quality assurance stereotactic radiosurgery |
|
|