Accuracy of intraocular pressure measurements in new zealand white rabbits |
| |
Authors: | Lim K Sheng Wickremasinghe Sanjeewa S Cordeiro M Francesca Bunce Catey Khaw Peng T |
| |
Affiliation: | Glaucoma Service, Moorfields Eye Hospital, London, UK. shenglim@gmail.com |
| |
Abstract: | PURPOSE: Rabbits are commonly used for the evaluation of drugs and surgery to lower intraocular pressure (IOP). The accuracy of intraocular pressure measurement is therefore critical in the analysis of data and subsequent extrapolation to humans. The purpose of this study was to establish the most reliable technique, from several currently available methods, of measuring IOP in New Zealand White rabbits. METHODS: The IOPs of 11 healthy New Zealand White rabbits were measured with a fluid-filled pressure transducer system that was connected to the anterior chamber of each animal so that the IOP could be varied by altering the height of a bottle of balanced salt solution. Intraocular pressures were recorded over a range of 50 to 0 mm Hg by the transducer system, and comparative measurements at the same pressures were performed with the Tonopen XL (Mentor, Norwell, MA), the Perkins handheld applanation tonometer (Clement Clarke, Harlow, UK), and the Ocular Blood Flow (OBF) pneumatonometer (OBF Laboratories, Wiltshire, UK). RESULTS: All three tonometers underestimated the true IOP, especially at higher pressures. Although the mean difference in actual and tonometric IOP was least in the Tonopen XL, there was a high degree of variability through the entire range of IOPs measured. The Perkins tonometer, although slightly less accurate than the Tonopen XL, was more consistent in measurement. The OBF pneumatonometer was the least-reliable estimate of IOP. CONCLUSIONS: All tonometers underestimate IOP, with increasing inaccuracy at higher pressures. The Perkins applanation tonometer was the most reliable for measuring IOP in the range of 0 to 50 mm Hg in New Zealand White rabbits. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|