Determinants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environment
Authors:Harikrishna Tandri MD   Menekhem M. Zviman PhD   Steven R. Wedan MS   Thomas Lloyd MS   Ronald D. Berger MD   PhD   FHRS  Henry Halperin MD   MA   FHRS
Affiliation:Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA. htandri1@jhmi.edu
Abstract:BACKGROUND: The determinants of low-frequency-induced current by magnetic resonance imaging (MRI) gradient fields in a pacemaker lead system are largely unknown. OBJECTIVE: The purpose of this study was to determine the magnitude of MRI low-frequency-induced current in an implanted pacemaker lead system and to investigate in vivo determinants of low-frequency-induced current in an animal model. METHODS: Six mongrel dogs underwent conventional single-chamber pacemaker implantation with a current recorder connected in series. Pulse generator (PG) was programmed to VOO 120 bpm with subthreshold output. MRI was performed in a 1.5-T scanner. Low-frequency-induced current was recorded during unipolar pacing, bipolar pacing, and bipolar pacing with the PG case electrically isolated from the pocket. In each mode, low-frequency-induced current was recorded with and without a large loop of additional lead connected in series. RESULTS: With a conventional implant, low-frequency-induced current was < or =0.5 mA in all three pacing modes. With five external loops, the magnitude of low-frequency-induced current increased to >30 mA, with consistent myocardial capture in unipolar and bipolar pacing. However, in bipolar pacing with the PG electrically isolated from the pocket, low-frequency-induced current decreased to <0.5 mA with no myocardial capture even with additional looped leads. CONCLUSION: Under conventional implant conditions, the magnitude of low-frequency-induced current is <0.5 mA and is unlikely to cause myocardial capture; however, arrhythmia induction cannot be excluded. With sufficient increase in effective loop area (additional looped leads), direct myocardial capture by the low-frequency-induced current is possible. In this study, breaking the return pathway by electrically isolating the PG case from the circuit abolished low-frequency-induced current.
Keywords:Pacemaker   Magnetic resonance imaging   Induced current   Interaction   Determinants
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号