多方法检测肱-踝动脉脉搏波传导速度的比较研究
引用本文:任亚娟,肖沪生,徐芳,银浩强,高东雯,王奇. 多方法检测肱-踝动脉脉搏波传导速度的比较研究[J]. 中国中西医结合影像学杂志, 2014, 0(2): 116-118
作者姓名:任亚娟  肖沪生  徐芳  银浩强  高东雯  王奇
作者单位:上海中医药大学附属龙华医院超声科,上海200032
基金项目:上海市教委科研课题(shzy020).
摘    要:目的:比较分析同一组受试者运用多种方法检测获得的肱-踝动脉段脉搏波传导速度(brachial ankle pulse wave velocity,baPWV),探讨将超声技术引入脉搏波传导速度检测的可行性及准确性。方法:纳入30例正常人,分别行日本Colin动脉硬化仪检测、联合多普勒超声及脉搏波传感器同步检测、仅多普勒超声检测baPWV,并将结果进行比较分析。结果:动脉硬化仪检测得到的baPWV为(1 118.20±176.42)cm/s;将脉搏波传感器放置于肱动脉、多普勒探头放置于胫后动脉同步检测的baPWV为(1 135.74±246.26)cm/s;仅用多普勒超声分2次分别将探头放置于肱动脉及胫后动脉时测得的baPWV为(1 048.96±117.24)cm/s。多普勒超声及脉搏波传感器联合应用与动脉硬化仪检测得到的baPWV呈显著正相关(r=0.549,P0.01)。多普勒超声与动脉硬化仪检测得到的baPWV无相关性(r=0.277,P0.05)。结论:联合运用超声多普勒与脉搏波传感器为PWV的检测提供了新途径,测量数据准确度较高;仅运用多普勒超声分2次检测PWV的方法误差较大。

关 键 词:超声检查  脉搏波传导速度  对比研究

Comparative study on the detection of brachial ankle pulse wave velocity by multi-method
Affiliation:REN Ya-juan, XIAO Hu-sheng, XU Fang, et al.( Department of Ultrasound, Longhua Hospital, Shanghai University of TCM, Shanghai, 200032, China.)
Abstract:Objective:To explore the feasibility and accuracy of detecting the brachial -ankle pulse wave velocity (baPWV ) by ultrasonography technology through the comparative analysis of the results of brachial ankle pulse wave velocity detected by multi-method in the same group of subjects .Methods:Thirty cases of healthy people were enrolled .Three kinds of methods (①Joint using of Doppler probe and pulse wave sensor simultaneously .②Using Doppler probe only .③Using Colin arteriosclerosis detecting device) were applied to detect baPWV .Then the results were compared and analyzed .Results:The baPWV (brachial pulse wave-tibial Doppler) was (1135 .74 ± 246 .26) ,when the pulse wave sensor was placed in the brachial artery and the Doppler probe was placed in the posterior tibial artery simultaneously .The baPWV (Doppler) was (1048 .96 ± 117 .24) ,when the Doppler probe was placed in the brachial artery and the posterior tibial artery in two times respectively .The baPWV(arteri-osclerosis detecting device) was (1118 .20 ± 176 .42) ,which was detected by Colin arteriosclerosis detecting device .There were positive correlations and no statistical significance between the test result by arteriosclerosis detecting device and the test result when the ultrasonic Doppler probe was placed on the left posterior tibial artery while the pulse wave sensor was placed on the left brachial artery (r=0 .549 ,P〈0 .01) .There were no correlations (r=0 .277 ,P〉0 .05) and obviously statistical significance between baPWV(Doppler) and baPWV (arteriosclerosis detecting device) .Conclusions :Joint applying the ultrasonic probe and the pulse wave sensor provides a new method for the detection of pulse wave velocity .This method is with high accuracy and worth to be used in clinical area and to be further spread .Detecting pulse wave velocity by only using Doppler probe has more error rate .
Keywords:Ultrasonography  Brachial-ankle pulse wave velocity  Comparative study