首页 | 本学科首页   官方微博 | 高级检索  
     


A new weighting scheme for cone-beam helical CT to reduce the image noise
Authors:Taguchi Katsuyuki  Chiang Be-Shan S  Silver Michael D
Affiliation:Application & Research, CT Systems Development Department, Toshiba Medical Systems Corporation, 1385 Shimoishigami, Otawara, Tochigi 324-8550, Japan. ktaguchi@aol.com
Abstract:Reducing the patient dose while keeping the image noise at the same level is desired for x-ray CT examinations. In order to achieve the goal, we propose a new weighting scheme taking the validity of the data and redundant data samples into account. The method is evaluated with a new generalized version of the Feldkamp helical reconstruction algorithm. It allows us to enlarge the projection angular range used in reconstruction, and thus, to reduce the image noise by increasing the detector utilization rate to 100% without sacrificing the image quality or z-resolution. This concept can be adapted to other exact or approximate algorithms as far as they use redundant data samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号