Failed transcatheter creation of fenestration in a Fontan patient with original fenestration closed by AngelWings septal occluder: a case report |
| |
Authors: | Ohno Atsuko Hirotaka Ishido Kobayashi Toshiki Asano Haruhiko Kyo Shunei Senzaki Hideaki |
| |
Affiliation: | Department of Pediatric Cardiology, International Medical Center, Saitama Medical University, Saitama. |
| |
Abstract: | The fenestrated Fontan procedure has helped to reduce the mortality and morbidity of pediatric patients who underwent the Fontan operation. We treated a patient (7-year-old, female) who had developed heart failure 1 year after fenestration closure using an AngelWings device, although the patient met the proposed criteria for test occlusion. Hemodynamic evaluation using the ventricular pressure-area relationship revealed that the patient had both systolic and diastolic dysfunction, which were difficult to recognize by conventional hemodynamic indices. We then tried to percutaneously create a fenestration using a Brockenbrough needle, but this procedure failed. The AngelWings device, which may have been covered by pseudointima (composed of fibroelastic tissue), was hard enough to break the Brockenbrough needle. Re-creating a fenestration after device closure in patients after the fenestrated Fontan operation appears to be difficult, so more accurate methods to predict Fontan physiology after fenestration closure are needed. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|