首页 | 本学科首页   官方微博 | 高级检索  
     

可植入微型轴流血泵溶血实验研究
引用本文:王玉璇,张杰民,施超,刘晓程. 可植入微型轴流血泵溶血实验研究[J]. 山西医科大学学报, 2010, 41(2): 133-136
作者姓名:王玉璇  张杰民  施超  刘晓程
作者单位:1. 中国医学科学院,北京协和医学院,北京100730
2. 泰达国际心血管病医院辅助循环研究室
3. 中国医学科学院,北京协和医学院,北京100730;泰达国际心血管病医院辅助循环研究室
基金项目:天津市科技支撑计划重点项目 
摘    要:目的通过自制轴流血泵进行体外、体内溶血实验,探讨血泵的溶血特性。方法轴流血泵置于充满新鲜羊血的模拟循环通路中,在输出流量为5L/min,后负荷压力为100mmHg的条件下,检测血泵体外溶血水平。在3只健康绵羊体内植入血泵进行在体辅助实验,并稳定运转24h,测定其在体溶血水平。结果可植入微型轴流血泵的体外溶血指数(normal index of haematolysis,NIH)为(0.01644±0.00018)g/100L。在体实验中,血泵运行无机械故障,循环辅助后血浆游离血红蛋白(plasma free hemoglobin,FHb)及乳酸脱氢酶(lactate dehydrogenase,LDH)呈现上升趋势,平均最大值分别达(7.93±0.71)g/L、(622.3±22.5)U/L。结论血泵溶血性能在可接受范围内,接近于部分国外三代血泵溶血水平,初步具备进入动物辅助实验研究水平。

关 键 词:心室辅助装置  轴流血泵  溶血实验

Hemolysis test about a self-made miniature implantable axial flow blood pump in vitro and in vivo
WANG Yu-xuan,ZHANG Jie-min,SHI Chao,LIU Xiao-cheng. Hemolysis test about a self-made miniature implantable axial flow blood pump in vitro and in vivo[J]. Journal of Shanxi Medical University, 2010, 41(2): 133-136
Authors:WANG Yu-xuan  ZHANG Jie-min  SHI Chao  LIU Xiao-cheng
Affiliation:WANG Yu-xuan1,ZHANG Jie-min2,SHI Chao1,LIU Xiao-cheng1,2 (1Chinese Academy of Medical Sciences,Peking Union Medical College,Beijing 100730,China,2Laboratory of Mechanical Circulatory Support,TEDA International Cardiovascular Hospital)
Abstract:Objective To explore the hemolytic characteristics of a self-made implantable axial flow blood pump in vitro and in vivo. Methods Blood pumps were connected with the pump in a mock circulatory loop,filled with fresh sheep blood,and the pump generated 5 L/min flow against a head pressure of 100 mmHg.The vitro hemolytic properties were evaluated by determining the level of plasma free hemoglobin(FHb)at defined intervals.Blood pump was implanted in 3 healthy sheep for in vivo test.After pumping,FHb,lactate deh...
Keywords:ventricular assist devices  miniature axial blood pump  hemolysis test  