A design study for a 'spiral staircase' ionization chamber for the quality control of electron beams |
| |
Authors: | Jenkins N H Barnes P J Bonnett D E |
| |
Affiliation: | Department of Medical Physics, Kent Oncology Centre, Maidstone and Tunbridge, Wells NHS Trust, Maidstone, Kent, ME16 9QQ, UK. |
| |
Abstract: | In order to verify that the energies of electron beams used for external beam therapy remain constant, IPEM 81 recommends a constancy check based on the ratio of ionization chamber measurements at two depths along the central axis. Such measurements for a range of electron energies can be a time consuming process. The purpose of this study was to design a device that would use several ion chambers simultaneously to measure electron depth dose curves, and hence the electron energy. A design was developed for a device consisting of ten independent ionization chambers, shaped and arranged in a solid phantom like the steps of a spiral staircase, the axis of the staircase being coincident with the axis of the electron beam. Measurements were carried out to test the design of individual chambers and to optimize the radius of the spiral and both the depth intervals and the lateral spacing between adjacent chambers. For ranges of electron energy from 6-12 MeV and 12-20 MeV the radii of the spirals needed were found to be 36.5 mm and 30.9 mm, the angular separations between edges of the chambers were 52 degrees and 30 degrees and chamber depths were found to be 10, 15, 20, 25, 30, 35, 40, 45, 50, 55 mm and 20, 40, 45, 50, 55, 60, 65, 70, 75, 80 mm, respectively. |
| |