Improvement in image quality of noncontrast head images in multidetector-row CT by volume helical scanning with a three-dimensional denoising filter
Authors:Sasaki Tadashi  Sasaki Makoto  Hanari Takao  Gakumazawa Hiroshi  Noshi Yasuhiro  Okumura Miwa
Affiliation:(1) Center for Radiological Sciences, Iwate Medical University Hospital, 19-1 Uchimaru, Morioka 020-8505, Japan;(2) Department of Radiology, Iwate Medical University, Morioka, Japan;(3) Toshiba Medical Systems Corporation, Otawara, Japan
Abstract:Purpose The aim of this study was to improve the contrast-to-noise ratio on noncontrast head computed tomography (CT) images, which are crucial for assessing patients with acute ischemic stroke. We applied a technique combining volume helical scanning with a three-dimensional (3D) denoising filter. Materials and methods We scanned phantoms for low-contrast resolutions and helical/cone-beam artifacts as well as stroke patients using a 16-row multidetector-row CT (MDCT) unit. Volume helical scans with 1-mm collimation and nonhelical scans with 8-mm thickness were performed. From the 1-mm thick volume data, 8-mm thick contiguous images were generated before and after applying a 3D denoising filter. Results On images stacked from volume data, the contrast-to-noise ratio was significantly improved by the 3D denoising filter and was nearly the same as that on nonhelical images. On stacked volume images, artifacts due to the cone beam and the helical scan were increased with larger helical pitches, but bone-related streak artifacts in the posterior fossa and underneath the calvarium were reduced when compared with nonhelical images. Conclusion Volume helical scan with a 3D denoising filter effectively improves image quality in noncontrast head MDCT images.
Keywords:Multidetector-row CT  Noncontrast head CT  Denoising filter  Helical scan  Low-contrast resolution
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号