Efficacy of forced-air warming systems with full body blankets
Authors:Bräuer Anselm  J M Michael  Steinmetz Nicolas  Lorenz Nadja  Perl Thorsten  Weyland Wolfgang  Quintel Michael
Institution:1. Department of Anesthesiology, Emergency and Intensive Care Medicine, University of G?ttingen, Germany
3. Department of Anesthesia, Montreal General Hospital and McGill University, Montreal, Quebec, Canada
4. Department of Anesthesia and Intensive Care Medicine, Evangelisches Bethesda-Krankenhaus Essen, Germany
Abstract:PURPOSE: Postoperative hypothermia after cardiac surgery is still a common problem often treated with forced-air warming. This study was conducted to determine the heat transfer efficacy of 11 forced-air warming systems with full body blankets on a validated copper manikin. METHODS: The following systems were tested: 1) Bair Hugger 505; 2) Bair Hugger 750; 3) Life-Air 1000 S; 4) Snuggle Warm; 5) Thermacare; 6) Thermacare with reusable Optisan blanket; 7) WarmAir; 8) Warm-Gard; 9) Warm-Gard and reusable blanket; 10) WarmTouch; and 11) WarmTouch and reusable blanket. Heat transfer of forced-air warmers can be described as follows: Q = h x DeltaT x A. Where Q = heat flux (W), h = heat exchange coefficient (W x m-2 x degrees C-1), DeltaT = temperature gradient between blanket and manikin surface (degrees C), A = covered area (m2). Heat flux per unit area and surface temperature were measured with 16 heat flux transducers. Blanket temperature was measured using 16 thermocouples. The temperature gradient between blanket and surface (DeltaT) was varied and h was determined by linear regression analysis. Mean DeltaT was determined for surface temperatures between 32 degrees C and 38 degrees C. The covered area was estimated to be 1.21 m2. RESULTS: For the 11 devices, heat transfers of 30.7 W to 77.3 W were observed for surface temperatures of 32 degrees C, and between -8.8 W to 29.6 W for surface temperatures of 38 degrees C. CONCLUSION: There are clinically relevant differences between the tested forced-air warming systems with full body blankets. Several systems were unable to transfer heat to the manikin at a surface temperature of 38 degrees C.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号