Effect of hemolysis on cardiac troponin T determination by the Elecsys 2010 immunoanalyzer
Authors:Lyon Martha E  Ball Cindy L  Krause Richard D  Slotsve George A  Lyon Andrew W
Affiliation:Department of Pathology and Laboratory Medicine, University of Calgary, Calgary, AB, Canada. martha.lyon@cls.ab.ca
Abstract:OBJECTIVE: To evaluate the effect of hemolysis on the measurement of cardiac troponin T (cTnT) using the Elecsys 2010 immunoanalyzer. METHODS: cTnT concentrations were measured using the Elecsys 2010. Interference studies were conducted by mixing serum of known cTnT concentration with either hemolysates or serum to which purified hemoglobin (Hgb) had been added. Hemolysates were prepared by mechanical disruption. PROBIT analysis was conducted to determine the probability of a cTnT result of >0.1 microg/l being decreased to <0.1 microg/l due to hemoglobin interference. RESULTS: Purified hemoglobin as well as serum-hemolysates mediated a concentration-dependent inhibition of cTnT determination with the Elecsys 2010. With every 1 g/l increase in hemoglobin, the probability of cTnT >0.1 is decreased by 2.5%. CONCLUSIONS: A concentration-dependent negative bias in the measurement of cTnT is associated with increasing hemoglobin concentrations in serum.
Keywords:Interference   Hemolysis   Troponin T   Hemoglobin