首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of three semiautomatic methods for determination of left ventricular ejection fraction from gated cardiac blood pool images
Authors:John Bingham  Dr. Robert Okada  Kenneth McKusick  Charles Boucher  Edward Tarolli  Nathaniel Alpert  William Strauss
Affiliation:1. Division of Nuclear Medicine, Massachusetts General Hospital, USA
2. the Cardiac Unit, Massachusetts General Hospital, USA
Abstract:A number of computer programs exist for the determination of ejection fraction from gated blood pool images. Three commercially available algorithms were compared on the same radionuclide data from 50 catheterized patients. The methods used were: (1) simple thresholding, (2) second derivative edge determination, and (3) a combination of these two. There was good correlation with catheterization ejection fraction for all three methods using smoothed data [r=0.781, 0.77, 0.84 respectively (P<0.01)]. However, interobserver variance was significantly decreased (P<0.01) by use of the second derivative method when compared with the other two methods and catheterization. Thus, if accuracy and low interobserver variance are required, then a combination of a second derivative and thresholding method is preferable.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号