Comparison of three semiautomatic methods for determination of left ventricular ejection fraction from gated cardiac blood pool images |
| |
Authors: | John Bingham Dr. Robert Okada Kenneth McKusick Charles Boucher Edward Tarolli Nathaniel Alpert William Strauss |
| |
Affiliation: | 1. Division of Nuclear Medicine, Massachusetts General Hospital, USA 2. the Cardiac Unit, Massachusetts General Hospital, USA
|
| |
Abstract: | A number of computer programs exist for the determination of ejection fraction from gated blood pool images. Three commercially available algorithms were compared on the same radionuclide data from 50 catheterized patients. The methods used were: (1) simple thresholding, (2) second derivative edge determination, and (3) a combination of these two. There was good correlation with catheterization ejection fraction for all three methods using smoothed data [r=0.781, 0.77, 0.84 respectively (P<0.01)]. However, interobserver variance was significantly decreased (P<0.01) by use of the second derivative method when compared with the other two methods and catheterization. Thus, if accuracy and low interobserver variance are required, then a combination of a second derivative and thresholding method is preferable. |
| |