Real-Time Contrast Echo Assessment of Myocardial Perfusion at Low Emission Power: First Experimental and Clinical Results Using Power Pulse Inversion Imaging
Authors:Tiemann Klaus  Lohmeier Stefan  Kuntz Stefanie  Köster Jörg  Pohl Christoph  Burns Peter  R Thomas  Nanda Navin C.  Lüderitz Berndt  Becher Harald
Affiliation:Department of Cardiology, University of Bonn, Bonn, Germany;Department of Medical Biophysics, University of Toronto, Toronto, Canada;University of Nebraska Medical Center, Omaha, Nebraska;Heart Station, Division of Cardiology, University of Alabama at Birmingham, Birmingham, Alabama
Abstract:Power pulse inversion (PPI) has been developed for echocontrast specific imaging in order to reduce destruction of microbubbles. The purpose of this study was to evaluate PPI for real-time contrast echocardiography. Therefore, in vitro studies in a physiological flow-phantom and clinical examinations in patients with coronary artery disease were performed. The in vitro rersults of this study indicate that PPI allows real-time imaging at low emission power and is almost nondestructive to contrast microbubbles of Definity. At this low emission power a strong linear relationship between the dosage of the contrast agent and the resulting PPI signal intensity was found (R = 0.998, p < 0.001). In the clinical examinations real-time imaging using low mechanical index PPI resulted in strong myocardial signals and a complete filling of the cavities indicating absence of bubble destruction. Most striking was the ability of PPI to display myocardial thickening and wall motion simultaneously with the assessment of myocardial contrast replenishment following ultrasound induced bubble destruction by high power frames. We conclude that PPI allows nondestructive contrast imaging both in experimental and clinical settings. Therefore, real-time imaging of myocardial perfusion and real-time assessment of contrast replenishment following ultrasound induced destruction of microbubbles is feasible. Moreover, PPI allows simultaneous assessment of perfusion and myocardial function.
Keywords:power pulse inversion imaging    contrast echocardiography    myocardial perfusion    low emission power
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号