仪器维护与技术培训一体化在流式细胞仪管理中的应用
引用本文:付伟超,安崇佑,任彦松,孙静,文亚男,张森,于文颖,梁昊岳.仪器维护与技术培训一体化在流式细胞仪管理中的应用[J].医疗卫生装备,2021(2):66-72,105.
作者姓名:付伟超  安崇佑  任彦松  孙静  文亚男  张森  于文颖  梁昊岳
作者单位:中国医学科学院血液病医院;天津市中心妇产科医院
基金项目:北京协和医学院2019年度校级研究生教育教学改革项目(10023201901402);中国医学科学院血液病医院(中国医学科学院血液学研究所)第二批管理课题(GL1911)。
摘    要:以BD CantoⅡ流式细胞仪为例,介绍了检测正常人外周血na觙ve细胞和静息态T细胞的标准方法,重点演示了仪器校准、电压调节、补偿调节等实验模板的建立流程;以BD Influx流式细胞仪为例,介绍了流式细胞仪的结构与功能特点以及平台多色流式细胞术技术培训中的相关内容;列举了BD Influx流式细胞仪的常见故障、引起...

关 键 词:BD  Influx流式细胞仪  BD  Canto  Ⅱ流式细胞仪  仪器维护  技术培训  故障分析

Application of integrated instrument maintenance and technical training in flow cytometer management
FU Wei-chao,AN Chong-you,REN Yan-song,SUN Jing,WEN Ya-nan,ZHANG Sen,YU Wen-ying,LIANG Hao-yue.Application of integrated instrument maintenance and technical training in flow cytometer management[J].Chinese Medical Equipment Journal,2021(2):66-72,105.
Authors:FU Wei-chao  AN Chong-you  REN Yan-song  SUN Jing  WEN Ya-nan  ZHANG Sen  YU Wen-ying  LIANG Hao-yue
Institution:(State Key Laboratory of Experimental Hematology,National Clinical Research Center for Blood Diseases,Institute of Hematology&Blood Diseases Hospital,Chinese Academy of Medical Sciences&Peking Union Medical College,Tianjin 300020,China;Tianjin Central Hospital of Gynecology Obstetrics,Tianjin 300052,China)
Abstract:Using BD CantoⅡflow cytometer as an example,the standard method for detecting native cells and resting T cells in normal human peripheral blood was introduced,and the process of establishing experimental templates such as instrument calibration,voltage adjustment and compensation adjustment was demonstrated;taking BD Influx flow cytometer as an example,the structure and functional features of the flow cytometer were described along with the technical training for the platform multi-colour flow cytometry.The common faults of BD Influx flow cytometers,their causes and treatment methods were listed,and the integrated management mode of routine maintenance and special technical training of the flow cytometers were discussed,aiming to provide references for relevant instrument managers and users as well as research institutes to improve the management and service of the flow cytometer.
Keywords:BD Influx flow Cytometer  BD CantoⅡflow cytometer  instrument maintenance  technical training  fault analysis
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏