Development of a miniature motor-driven pulsatile LVAD driven by a fuzzy controller
Authors:Eiji Okamoto PhD  Tsutomu Makino PhD  Shuji Tanaka MS  Takahiko Yasuda MS  Yuta Akasaka BS  Makiko Tani BS  Yusuke Inoue MS  Ayumu Mitoh PhD  Yoshinori Mitamura PhD
Affiliation:(1) Department of Information Science, School of Engineering, Hokkaido Tokai University, 5-1-1-1 Minami-sawa, Minami-ku, Sapporo 005-8601, Japan;(2) Graduate School of Information Science and Technology, Hokkaido University, Sapporo, Japan;(3) Tomakomai College of Technology, Tomakomai, Japan
Abstract:We have been developing a small, lightweight motor-driven pulsatile left ventricular assist device (LVAD) with a ball screw. The motor-driven LVAD consists of a brushless DC motor and a ball screw. The attractive magnetic force between Nd–Fe–B magnets (with a diameter of 5 mm and a thickness of 1.5 mm) mounted in holes in a silicone rubber sheet (thickness 2 mm) and an iron plate adhered onto the a diaphragm of the blood pump can provide optimum active blood filling during the pump filling phase. The LVAD has a stroke volume of 55 ml and an overall volume of 285 ml; it weighs 360 g. The controller mainly consists of a fuzzy logic position and velocity controller to apply doctors' and engineers' knowledge to control the LVAD. Each unit of the controller consists of a functionally independent program module for easy improvement of the controller's performance. The LVAD was evaluated in in vitro experiments using a mock circulation. A maximum pump outflow of 5.1 l/min was obtained at a drive rate of 95 bpm against an afterload of 95 mmHg, and active filling using the attractive magnetic force provided a pump output of 3.6 l/min at a drive rate of 75 bpm under a preload of 0 mmHg. The operating efficiency of the LVAD was measured at between 8% and 10.5%. While the LVAD can provide adequate pump outflow for cardiac assistance, further upgrading of the software and improvement of the blood pump are required to improve pump performance and efficiency.
Keywords:Artificial heart  LVAD  Ball-screw  Fuzzy  Active filling
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号