首页 | 本学科首页   官方微博 | 高级检索  
     


Quantitative analysis of magnetic resonance imaging susceptibility artifacts caused by neurosurgical biomaterials: comparison of 0.5, 1.5, and 3.0 Tesla magnetic fields
Authors:Matsuura Hideki  Inoue Takashi  Ogasawara Kuniaki  Sasaki Makoto  Konno Hiromu  Kuzu Yasutaka  Nishimoto Hideaki  Ogawa Akira
Affiliation:Department of Neurosurgery, Iwate Medical University School of Medicine, Morioka, Iwate, Japan.
Abstract:Magnetic resonance (MR) imaging is an important diagnostic tool for neurosurgical diseases but susceptibility artifacts caused by biomaterial instrumentation frequently causes difficulty in visualizing postoperative changes. The susceptibility artifacts caused by neurosurgical biomaterials were compared quantitatively by 0.5, 1.5, and 3.0 Tesla MR imaging. MR imaging of uniform size and shape of pieces ceramic (zirconia), pure titanium, titanium alloy, and cobalt-based alloy was performed at 0.5, 1.5, and 3.0 Tesla. A linear region of interest was defined across the center of the biomaterial in the transverse direction, and the susceptibility artifact diameter was calculated. Susceptibility artifacts developed around all biomaterials at all magnetic field strengths. The artifact diameters caused by pure titanium, titanium alloy, and cobalt-based alloy increased in the order of 0.5, 1.5, to 3.0 Tesla magnetic fields. The artifact diameter of ceramic was not influenced by magnetic field strength, and was the smallest of all biomaterials at all magnetic field strengths. The artifacts caused by biomaterials except ceramic increase with the magnetic field strength. Ceramic instrumentation will minimize artifacts in all magnetic fields.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号