首页 | 本学科首页   官方微博 | 高级检索  
检索        


Accuracy of computer-navigated total hip arthroplasty
Authors:Snyder Garrett M  Lozano Calderón Santiago A  Lucas Paul A  Russinoff Scott
Institution:Department of Orthopaedic Surgery, Westchester Medical Center, Valhalla, New York, USA.
Abstract:Proper acetabular cup orientation is essential in total hip arthroplasty. The purpose of this study was to evaluate the accuracy of a particular imageless computer navigation system in determining cup position. Thirty-nine computer-navigated total hip arthroplasty intraoperative measurements of cup abduction and anteversion were compared with those from follow-up radiographs. Sensitivity, specificity, accuracy, prevalence-adjusted positive value (PPV), and negative predictive value were calculated for both navigation and radiographs. Navigation measurements had high specificity and PPV when assessing cup abduction and anteversion (specificity >90%, PPV >94%). In contrast, the system was not very effective in detecting suboptimal cup position (sensitivity abduction, 50%; anteversion, 33%). Intraoperative navigation readings in the safe zone have high probability of indicating correct placement. However, confirmation of suboptimal cup position intraoperatively requires additional diagnostic methods.
Keywords:computer navigation  total hip arthroplasty  acetabular cup position  predictive values  accuracy
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏