首页 | 本学科首页   官方微博 | 高级检索  
     


Relationship between image quality and ultrasound exposure level in diagnostic US devices
Authors:Harris, GR   Stewart, HF   Leo, FP   Sanders, RC
Affiliation:Center for Devices and Radiological Health, Food and Drug Administration, Rockville, MD 20857.
Abstract:Ultrasonographic B-mode images were obtained at various exposure levels with three real-time diagnostic scanners. Adult human and tissue-equivalent phantom images were compared in terms of diagnostic content and depth of penetration. For the exposure level settings used, spatial-peak pulse-average intensities ranged from approximately 10 to 500 W/cm2. At the 3.50-3.75-MHz nominal frequencies used in the study, images of the human abdomen showed little discernible change in quality with varying exposure levels. However, phantom tests confirmed that depth of penetration is a function of exposure level. The results suggest that a judicious use of exposure level and receiver gain controls can be a practical means for minimizing patient exposure to ultrasound without sacrifice of diagnostic effectiveness.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号