Variability in Telavancin Cross-Reactivity among Vancomycin Immunoassays
Authors:Kevin W McConeghy  Siyun Liao  Douglas Clark  Philip Worboys  Steven L Barriere  Keith A Rodvold
Institution:aUniversity of Illinois at Chicago, College of Pharmacy, Chicago, Illinois, USA;bTheravance Biopharma US, South San Francisco, California, USA;cTheravance, Inc., South San Francisco, California, USA;dUniversity of Illinois at Chicago, College of Medicine, Chicago, Illinois, USA
Abstract:Telavancin is a semisynthetic lipoglycopeptide with a dual mechanism of action against Gram-positive pathogens. Two brief reports have suggested potential cross-reactivity of telavancin with the vancomycin particle-enhanced turbidometric immunoassay (PETIA). The purpose of this study was to evaluate several commercially available vancomycin immunoassays (fluorescence polarization FPIA], enzyme-multiplied immunoassays EMIT], PETIA, and chemiluminescent immunoassay CMIA]) for cross-reactivity with telavancin. Seven sites were selected to analyze serum samples for vancomycin. Each site received a set of samples (n = 18) which combined drug-free serum with telavancin, 7-OH telavancin metabolite, or vancomycin. Immunoassays demonstrating potential cross-reactivity were further evaluated by sending a duplicate sample set to multiple laboratories. Cross-reactivity was defined as the percent theoretical concentration (reported concentration/theoretical concentration × 100). No cross-reactivity was seen with FPIA or EMIT. Within the theoretical concentration range of 5 to 120 μg/ml of telavancin, the Synchron PETIA system reported vancomycin concentrations ranging from 4.7 to 54.2 μg/ml compared to vancomycin concentrations from 1.1 to 5.6 μg/ml for the Vista PETIA system. The Architect CMIA system reported vancomycin concentrations in the range of 0.27 to 0.97 μg/ml, whereas Advia Centaur XP CMIA reported vancomycin concentrations between 1.6 and 31.6 μg/ml. The Architect CMIA immunoassay had the lowest percent cross-reactivity (0.8 to 5.4%), while the Synchron PETIA immunoassay demonstrated the highest percent cross-reactivity (45.2 to 53.8%). Telavancin samples measured by liquid chromatography-mass spectroscopy were within 93.9 to 122% of theoretical concentrations. Vancomycin concentrations were not measured in any 7-OH telavancin-spiked sample. Vancomycin concentrations measured by liquid chromatography-mass spectroscopy were within 57.2 to 113% of theoretical concentrations. PETIA and CMIA measured vancomycin concentrations in telavancin-spiked samples. Significant variability in percent cross-reactivity was observed for each platform regardless of immunoassay method.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号