Continuous In-Line Monitoring of Oxygen Delivery to Control Artificial Heart Output
Authors:S Takatani  H Noda  H Takano  T Akutsu
Institution:Department of Biomedical Engineering and Artifical Organs, National Cardiovascular Center, Research Institute, Osaka, Japan.
Abstract:In order to evaluate the pump output control based on the oxygen delivery to peripheral tissues, arterial and mixed venous hemoglobin content (Hb]) and oxygen saturation (SO2) were continuously monitored in three biventricular bypass animals (3-, 6-, and 40-day experiments) with fibrillating ventricles. The specially developed oxygen sensors were mounted in the outflow ports of the artificial hearts to measure Hb] and SO2. One animal was exercised on the treadmill at 2.0 mile/h for 15 min with pump flows fixed to deliver oxygen of (a) above 13 cc/min/kg, (b) 10, and (c) 9. In (a), the mixed venous saturation (SvO2) dropped to approximately 25% with no increase in the blood lactate level. In (b) and (c), the SvO2 decreased to approximately 10-15% with increase in blood lactate levels from 4 to 10-30 mg/dl. Also, the recovery of the SvO2 in these groups following the termination of the exercise was slower in comparison to (a). The lower limit of the SvO2 level that would create oxygen debt situation in the peripheral tissues was approximately 25-30% for the exercise of 2.0 mile/h. The SvO2 reflects changes in respiratory status, pump output, hemoglobin level, and metabolism, and is thus a useful indicator to diagnose quickly the circulatory status as well as possibly to control the artificial heart output.
Keywords:SvO2  Blood lactate  Pump output control  Oxygen delivery  Optical sensor
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号