首页 | 本学科首页   官方微博 | 高级检索  
     

不同化学发光分析系统在皮质醇检测中的性能评价与比较
引用本文:陈翊,李颖能,陈焕辉,林涛,汤勇才. 不同化学发光分析系统在皮质醇检测中的性能评价与比较[J]. 国际检验医学杂志, 2016, 0(14). DOI: 10.3969/j.issn.1673-4130.2016.14.019
作者姓名:陈翊  李颖能  陈焕辉  林涛  汤勇才
作者单位:1. 广东省广州市妇女儿童医疗中心中心实验室 510120;2. 广东省广州市第一人民医院检验科 510180
基金项目:广东省广州市科技计划项目(2014Y2-00031)。
摘    要:目的比较贝克曼Access 2Immunoassay System化学发光免疫分析系统及罗氏COBAS E 411电化学发光免疫分析系统对皮质醇的分析性能与检测结果相关性。方法按照ISO15189中方法评估贝克曼及罗氏分析系统对皮质醇检测的精密度、功能灵敏度、参考范围、方法学、阳性符合率、阴性符合率。结果贝克曼与罗氏分析系统在皮质醇检测中精密度良好,功能灵敏度符合各自厂商设定的临床检测的下限,实验室自建参考范围后,将两分析系统对皮质醇的测定值进行线性回归分析,二者间相关性良好,线性回归系数r~2=0.977,阳性符合率、阴性符合率均为100%。结论两分析系统对于皮质醇的分析性能优异,均可满足临床检验需求。经过方法学比对,二者相关性良好,能够通过回归方程对二者间测定值进行引证。

关 键 词:皮质醇  化学发光免疫分析  电化学发光免疫分析  性能评价

Performance evaluation and comparison of different chemiluminescence immunoassay systems in detection of cortisol
Abstract:Objective To compare the analytical performance and results correlation of cortisol detection by the Beckman Access 2 Immunoassay System chemiluminescence system and the Roche COBAS E 411 electrochemical luminescence system .Methods The precision ,functional sensitivity ,reference range ,positive coincidence rate and negative coincidence rate of both Beckman Access 2 Immunoassay System and Roche COBAS E 411 systems for detecting cortisol were evaluated according to the method of ISO15198 .Results Both Beckman Access 2 Immunoassay System and Roche COBAS E 411 system exhibited better precision ,the functional sensitivity conformed to the lower limit of clinical detection set by the manufacturers ,after establishing the reference range of cortisol by the laboratory ,the cortisol values detected by these two systems were performed the linear regression analysis , the results indicated that the correlation between them was good ,the linear regression coefficient r2 =0 .977 ,and the positive and negative coincidence rates were 100% .Conclusion The two analytical systems have excellent performance for cortisol ,which all meet the clinical detection requirements .The methodological comparison shows good correlation between them ,and the detection re‐sults could be cited between these two systems by the regression equation .
Keywords:cortisol  chemoluminescence  electrochemical luminescence  performance evaluation
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号