首页 | 本学科首页   官方微博 | 高级检索  
     


The complementary value of Doppler-ultrasonography and ocular pneumoplethysmography according to Gee (OPG-Gee) in carotid obstructions
Authors:F Strik
Affiliation:Department of Neurology, Hospital HdK, Oldenzaal, The Netherlands.
Abstract:The results of the combination of CW-Doppler ultrasonography and OPG-Gee are described in 68 patients. The Doppler examination had a sensitivity of 92% and a specificity of 89%, the percentages for the OPG-Gee were 69 en 79. The combination of the Doppler examination with the OPG-Gee improves our positive predictive value from 83 to 94%, the negative predictive value from 95 to 98% in concordant results. OPG-Gee fails especially in low-graded (50-74%) stenoses. Since there is no correlation between the degree of the stenosis and the pathophysiological changes of the cerebral circulation, OPG-Gee is of complementary value for the evaluation of the hemodynamic significance of a stenosis, established with the Doppler method.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号