首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison and modification of two cerebral protection devices used for carotid angioplasty: in vitro experiment
Authors:Müller-Hülsbeck Stefan  Grimm Jan  Liess Carsten  Hedderich Jürgen  Bergmeyer Marcus  Heller Martin
Affiliation:Department of Radiology, University Hospital, Arnold-Heller-Strasse 9, 24105 Kiel, Germany. muehue@rad.uni-kiel.de
Abstract:The effectiveness of two basic cerebral protection devices designed for carotid angioplasty with and without additional aspiration techniques was compared in an in vitro model. During carotid angioplasty, embolization was simulated by injecting polyvinyl alcohol particles of different sizes into the model system. None of the tested devices, all of which were positioned in the internal carotid artery, was able to completely prevent embolization. In the internal carotid artery, the rate of particle capture did not vary among protection devices. However, embolization into the external carotid artery was more frequent with use of the GuideWire, as compared with that with use of the Angioguard.
Keywords:
本文献已被 PubMed 等数据库收录!
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号