新型电磁搏动式血泵的实验研究
引用本文:张磊,方旭晨,葛斌.新型电磁搏动式血泵的实验研究[J].医用生物力学,2019,34(2):173-178.
作者姓名:张磊  方旭晨  葛斌
作者单位:上海理工大学 医疗器械与食品学院,上海市杨浦区市东医院 危重医学科,上海理工大学 医疗器械与食品学院
基金项目:上海市大学生创新计划基金项目 (2016118)
摘    要:目的探究一种新型电磁搏动式血泵的动力输出性能及血液相容性。方法首先通过建立理论模型对该血泵驱动力进行分析,并基于该模型计算出满足条件的实验驱动电压。设计体外模拟循环实验,对新型血泵的输出流量和输出压力特性及血泵的体外溶血性能进行初步实验研究。结果实验测得当血泵后负荷为73.5 mmHg(9.78 kPa, 1 mmHg=0.133 kPa)、驱动电压达到35 V、搏动频率为75/min时,实际输出的流量为3.18 L/min,可以提供高压132 mmHg(17.56 kPa)、低压66 mmHg(8.78 kPa)、平均压力98 mmHg(13.03 kPa),体外实验标准溶血指数(normalized index of haematolysis, NIH)为(0.049 15±0.003 75) mg/dL。结论该新型搏动式血泵能够满足离体器官灌注和体外循环短期辅助的临床要求,对体外循环血泵的发展具有重要意义。

关 键 词:搏动泵    电磁驱动    溶血实验
收稿时间:2018/10/7 0:00:00
修稿时间:2018/11/10 0:00:00

Experimental Study on a Novel Electromagnetic Pulsatile Blood Pump
ZHANG Lei,FANG Xucheng and GE Bin.Experimental Study on a Novel Electromagnetic Pulsatile Blood Pump[J].Journal of Medical Biomechanics,2019,34(2):173-178.
Authors:ZHANG Lei  FANG Xucheng and GE Bin
Institution:School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology,Department of Critical Care Medicine, East City Hospital of Shanghai Yangpu District and School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology
Abstract:Objective To explore the power output and blood compatibility of a novel electromagnetic pulsatile blood pump. Methods First, a theoretical model was established to analyze the driving force of the blood pump, and the experimental driving voltage satisfying the conditions was calculated based on this model. Then, the output flow rate, output pressure characteristics and hemolysis performance of the new blood pump in vitro were preliminarily analyzed by simulated circulation experiments. Results The experimental result showed that when the pump load was 73-5 mmHg (9-78 kPa, 1 mmHg=0-133 kPa), the driving voltage was 35 V and the pulsation frequency was 75 beats/min, the flow rate of the pump was 3-18 L/min, producing high, low and average pressure of 132, 66, 98 mmHg (17-56, 8-78, 13-03 kPa), and the normalized index of haematolysis (NIH) in vitro was (0-049 15+0-003 75) mg/dL. Conclusions The new pulsatile blood pump can satisfy the clinical requirements for perfusion of isolated organs and short-term assistance of cardiopulmonary bypass, which is of great significance to the development of cardiopulmonary bypass blood pump.
Keywords:pulsatile pump  electromagnetic drive  hemolysis test