首页 | 本学科首页   官方微博 | 高级检索  
     

离子色谱法测定水中溴酸根离子含量的不确定度评价
引用本文:周小新,周虹. 离子色谱法测定水中溴酸根离子含量的不确定度评价[J]. 职业与健康, 2009, 25(17): 1827-1829
作者姓名:周小新  周虹
作者单位:江苏省无锡市疾病预防控制中心.214023
摘    要:目的进行离子色谱法测定水中溴酸根离子含量的不确定度分析。方法按照测定程序,建立数学模型,根据不确定度传播规律,计算出相对和绝对扩展不确定度。结果水样中溴酸根离子的浓度为0.0638mg/L时,其绝对扩展不确定度为±0.0031mg/L。结论该方法在实际工作中有较强的应用价值。

关 键 词:离子色谱法  溴酸根离子  不确定度

Uncertainty Evaluation of Bromate Ion Determination in Water by Ion Chromatography
ZHOU Xiao-xin,ZHOU Hong. Uncertainty Evaluation of Bromate Ion Determination in Water by Ion Chromatography[J]. Occupation and Health, 2009, 25(17): 1827-1829
Authors:ZHOU Xiao-xin  ZHOU Hong
Affiliation:( Wuxi Center for Disease Control and Prevention, Jiangsu, 214023,China)
Abstract:[ Objective] To analyze uncertainty of Bromate ion determination in water by ion chromatography. [ Methods ] According to procedures, mathematic model was established , then the relative and the absolute expanded uncertainties were calculated based on the spreading regulation of uncertainty. [ Results] When the concentration of Bromate in water arrived 0.0638 mg/L, the expand- ed uncertainty was ±0. 0031 mg/L. [ Conclusion]This method can produce large application value in practical work.
Keywords:Ion chromatography  Bromate ion  Uncertainty
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号