Pulse Conductance and Flow‐induced Hemolysis During Pulsatile Cardiopulmonary Bypass
Authors:Antoine P Simons  Pieter Wortel  Rob AT Van Kan  Frederik H Van Der Veen  Patrick W Weerwind  Jos G Maessen
Institution:Department of Cardiothoracic Surgery and Cardiovascular Research Institute Maastricht, Maastricht University Medical Center, Maastricht, The Netherlands
Abstract:In this study, the hypothesis was tested that a low‐resistant, high‐compliant oxygenator provides better pulse conductance and less hemolysis than a high‐resistant, low‐compliant oxygenator during pulsatile cardiopulmonary bypass. Forty adults undergoing coronary artery bypass surgery were randomly divided into two groups using either an oxygenator with a relatively low hydraulic resistance (Quadrox BE‐HMO 2000, Maquet Cardiopulmonary AG, Hirrlingen, Germany) or with a relatively high hydraulic resistance (Capiox SX18, Terumo Cardiovascular Systems, Tokyo, Japan). The phase shift between the flow signals measured at the inlet and outlet of the oxygenator was used to assess compliance. Pulse conductance in terms of pressure attenuation was calculated by dividing the outlet pulse pressure of the oxygenator by the inlet pulse pressure. A normalized index was used to assess hemolysis. The phase shifts in time of the flow pulses were 36 ± 6 ms in the low‐resistant (high‐compliant) oxygenator, and 14 ± 2 ms in the high‐resistant (low‐compliant) oxygenator group (P < 0.001). The low‐resistant, high‐compliant oxygenator provided 27% better pulse conductance compared with the high‐resistant, low‐compliant oxygenator (0.84 ± 0.02 and 0.66 ± 0.01, respectively, P < 0.001). Inlet pulse pressures were significantly higher (29%) in the high‐resistant, low‐compliant (Capiox) group than in the low‐resistant, high‐compliant (Quadrox) group (838 ± 38 mm Hg and 648 ± 25 mm Hg respectively, P < 0.001), but no significant difference in hemolysis was found. A low‐resistant, high‐compliant oxygenator provides better pulse conduction than a high‐resistant, low‐compliant oxygenator. However, the study data could not confirm the association of high pressures with increased hemolysis.
Keywords:Cardiopulmonary bypass  Oxygenator  Pulsatile flow  Compliance  Hemolysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号