Biomechanical evaluation of bone-cement augmented Proximal Femoral Nail Antirotation blades in a polyurethane foam model with low density |
| |
Authors: | Sermon A Boner V Schwieger K Boger A Boonen S Broos P Richards G Windolf M |
| |
Affiliation: | a Biomedical Services AO Research Institute, Clavadelerstrasse 8, 7270 Davos Platz, Switzerlandb Department of Traumatology, University Hospitals Gasthuisberg, Leuven, Belgiumc Research and Development Biomaterials, Synthes GmbH, Oberdorf, Switzerlandd Leuven University Center for Metabolic Bone Diseases and Division of Geriatric Medicine, Belgiume AO Research Institute, Davos, Switzerland |
| |
Abstract: | BackgroundHelically shaped cephalic implants have proven their benefit to provide an improved stabilization of unstable hip fractures. However, cut out ratios up to 3.6% still occur. This in vitro study evaluated the biomechanical performance of a novel cement augmentation technique of the Proximal Femoral Nail Antirotation in surrogate femora.MethodsFour study groups were formed out of 24 polyurethane foam specimens with low density. Proximal Femoral Nail Antirotation blades were implanted, either non-augmented, or augmented using 3 ml of injectable Polymethylmethacrylate bone-cement. The influence of implant mal-positioning was investigated by placing the blade either centered in the femoral head or off-centric in an anteroposterior direction. All specimens underwent cyclic loading under physiological conditions. Starting at 1000 N, the load was monotonically increased by 0.1 N/cycle until construct failure. Movement of the head was identified by means of optical motion tracking. Non-parametric test statistics were carried out on the cycles to failure, to compare between study groups.FindingsCompared to control samples; augmented samples showed a significantly increased number of cycles to failure (P = 0.012). In the groups with centric position of the Proximal Femoral Nail Antirotation blade, cement augmentation led to an increase in loading cycles of 225%. In the groups with off-centric positioning of the blade, this difference was even more accentuated (933%).InterpretationCement augmentation of the Proximal Femoral Nail Antirotation blade with small amounts of bone-cement for treatment of osteoporotic hip fractures clearly enhances fixation stability and carries high potential for clinical application. |
| |
Keywords: | Hip fracture Helical blade Cut out Augmentation PMMA Cement |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|