首页 | 本学科首页   官方微博 | 高级检索  
     


Influence of ambient humidity on the current delivered by air-vented ionization chambers revisited
Authors:Poirier Aurélie  Douysset Guilhem
Affiliation:Laboratoire National Henri Becquerel-LNE, CEA Saclay 91191 Gif-sur-Yvette, France.
Abstract:The influence of ambient humidity on the current delivered by a vented ionization chamber has been re-investigated. A Nucletron 077.091 well-type chamber together with a (192)Ir HDR brachytherapy source was enclosed in a climatic test chamber and the current was recorded for various humidity values. Great care has been taken for the design of the experimental setup in order to obtain reliable measurements of currents and humidity values inside the chamber active volume. A +/-0.35% linear variation of the measured currents has been observed over a common range of humidities. This result is larger than the expected variation. No formal explanation of such a discrepancy has been found yet, however the present results could lead to a set of recommendations.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号