首页 | 本学科首页   官方微博 | 高级检索  
     


Corneal hysteresis in patients with dry eye
Authors:Firat P G  Doganay S
Affiliation:Turgut Ozal Medical Center Department of Ophthalmology, Inonu University, Malatya, Turkey. penpe.firat@inonu.edu.tr
Abstract:

Purpose

To measure corneal hysteresis and intraocular pressure (IOP) in patients with dry eye and to compare with normal subjects.

Methods

This cross-sectional study consists of 70 eyes of 40 patients with dry eye (group 1) and 75 eyes of 40 normal subjects (group 2). Eyes were diagnosed as dry eye or normal according to the clinical symptoms, biomicroscopical evaluation, and Schirmer test. Corneal hysteresis (CH), corneal resistance factor (CRF), and cornea-compensated intraocular pressure (IOPcc) were measured by the Ocular Response Analyser (ORA).

Results

Mean CH values were 10.56±0.25 mm Hg and 10.34±0.26 mm Hg, mean CRF values were 10.75±0.28 mm Hg and 10.70±0.28 mm Hg, mean CCT values were 542±3.20 μm and 543±3.89 μm in group 1 and group 2, respectively. There was no statistically significant difference between the groups for these three parameters. IOPcc values measured with ORA were 15.73±0.36 mm Hg in group 1 and 16.60±0.33 mm Hg in group 2, and there was no statistically significant difference between the two groups.

Conclusions

Corneal biomechanical parameters such as CH and CRF are not influenced in dry eye. Also statistical difference was not found between the two groups according to CCT and IOPcc values.
Keywords:dry eye   corneal hysteresis   intraocular pressure   ocular response analyser
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号