A Case Report Demonstrating Spontaneous Change in Tachycardia Terminating Window |
| |
Authors: | KAREL den DULK PEDRO BRUGADA HEIN J.J. WELLENS |
| |
Affiliation: | Department of Cardiology, St. Annadal Hospital, Maastricht, The Netherlands |
| |
Abstract: | A patient who received a patient-activated pacemaker system for termination of paroxysmal, drug-resistant, ventricular tachycardia is described. During the pre-implant invasive electrophysiological (EP) study and the three post-implant non-invasive EP studies, tachycardias were easily and reproducibly terminated by two stimuli using the interval scanning mode. On the fourth day post-implant, this pacing mode no longer terminated tachycardia at rest. A new pacing mode using more stimuli was then found to terminate the tachycardia reproducibly under various physiological conditions. This case report demonstrated a spontaneous change in the tachycardia terminating window and the usefulness of versatility in antitachycardia pacemaker systems. |
| |
Keywords: | patient-activated pacing system externally-activated pacing system pacemaker termination of tachycardia change in tachycardia terminating window |
本文献已被 PubMed 等数据库收录! |
|