首页 | 本学科首页   官方微博 | 高级检索  
     


Hemodynamic effects of high frequency ventilation superimposed on intermittent positive pressure ventilation in dogs
Authors:T. Mogensen,M.D.,K. Eliasen,,L. Nielsen,,H. Gravesen,,H. Rygå  rd, J. B. Andersen
Affiliation:Department of Anaesthesiology and Radiology, University of Copenhagen, Herlev Hospital, Denmark.
Abstract:The hemodynamic effects of high frequency ventilation (HFV) superimposed on intermittent positive pressure ventilation (IPPV) in seven dogs before and after thrombin infusion were investigated. HFV was superimposed on a Servo 900 B ventilator by a Siemens Elema HFV prototype unit. Mean arterial blood pressure, heart rate, central venous pressure, pulmonary artery pressure, cardiac output, right and left ventricular pressures, pleural pressure, arterial blood gases, and right and left ventricular ejection fractions were recorded. Measurements were done during IPPV alone and during HFV superimposed on IPPV. The HFV frequencies were 5, 15, and 20 Hz at a constant minute volume of 5 1. When HFV was started, the IPPV minute volume was reduced to one third of the initial volume. No significant changes in the measured parameters were observed during the different ventilatory modes either before or after thrombin infusion which doubled the pulmonary vascular resistance. It is concluded that high frequency ventilation superimposed on IPPV might be a ventilatory mode that offers cardiovascular stability and reduces the risk of barotrauma.
Keywords:Artificial ventilation    hemodynamics    high frequency ventilation    thrombin infusion
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号