Prospective clinical evaluation of an improved fluorescence polarization assay for predicting fetal lung maturity
Authors:J F Talt  C A Foerder  E R Ashwood  T J Benedetti
Abstract:We performed a prospective clinical evaluation of our newly developed fluorescence polarization procedure to predict fetal lung maturity (Clin Chem 1986;32:248-54). Net fluorescence polarization was measured at 34 degrees C after a 6.5-min incubation of amniotic fluid with fluorophore. For the 26 cases of neonatal respiratory distress syndrome encountered in 196 deliveries, the net polarization exceeded 0.287 for 22 (85%) of these, and exceeded 0.260 for all 26. The specificity of the polarization assay equaled or exceeded the specificity of the lecithin/sphingomyelin ratio for all sensitivities greater than 70%. Neither assay was a good predictor of the clinical severity of respiratory distress. For a separate group of 21 amniotic fluid specimens clinically contaminated with blood or meconium, the discriminatory power of the polarization assay was decreased, but six of seven respiratory-distress cases still had polarization values greater than 0.260. We conclude that this fluorescence polarization assay is a better overall predictor of fetal lung maturity than is the lecithin/sphingomyelin ratio, and that polarization values less than 0.260 are associated with little risk of respiratory distress.