首页 | 本学科首页   官方微博 | 高级检索  
     


Detection of von Willebrand disorder and identification of qualitative von Willebrand factor defects. Direct comparison of commercial ELISA-based von Willebrand factor activity options
Authors:Favaloro E J
Affiliation:Department of Haematology, Westmead Hospital, Western Sydney Area Health Service, New South Wales, Australia.
Abstract:Two von Willebrand factor (vWF):collagen binding (activity) assay (CBA) kit methods are commercially available. A monoclonal antibody (MAB)-based enzyme-linked immunosorbent assay (ELISA) system reported to correlate with a standard vWF:ristocetin cofactor (RCof) assay is also commercially available. It is marketed as a vWF:Activity assay and is available in 2 assay version formats. In the present study, these 4 vWF-activity options were compared directly with in-house vWF:CBA ELISAs for their ability to detect von Willebrand disease (vWD) and identify qualitative vWF defects. The 2 MAB-based systems detected vWD but could not specifically identify qualitative vWF defects, although the recently modified Mark II kit was more effective for the latter compared with the original Mark I kit. All vWF:CBA methods, including in-house and commercial, also effectively detected vWD but differed in their ability to identify qualitative vWF defects. Effectiveness was highest using the in-house reference vWF:CBA (using a type I/III collagen mix product from equine tendon), the Gradipore vWF:CBA (also uses equine tendon-derived collagen), or the in-house vWF:CBA methods using type III human collagen at a relatively low concentration (1 or 3 micrograms/mL, without covalent linkage). The IMMUNO vWF:CBA seemed to be the least effective among the vWF:CBA methods for detection of qualitative vWF defects.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号