首页 | 本学科首页   官方微博 | 高级检索  
     


Agreement of Mixed Venous Carbon Dioxide Tension (PvCO2) and Transcutaneous Carbon Dioxide (PtCO2) Measurements in Ventilated Infants
Authors:Sinan Uslu  Ali Bulbul  Mesut Dursun  Umut Zubarioglu  Ebru Turkoglu  Omer Guran
Affiliation:1.Division of Neonatology, Sisli Hamidiye Etfal Educational and Research Hospital, Istanbul, Turkey
Abstract:

Background:

Noninvasive transcutaneous carbon dioxide monitoring has been shown to be accurate in infants and children, limited data are available to show the usefulness and limitations of partial transcutaneous carbon dioxide tension (PtCO2) value.

Objectives:

The current study prospectively determines the effectiveness and accuracy of PtCO2 measurements in newborns.

Materials and Methods:

Venous blood gas sampling and monitoring of the PtCO2 level (TCM TOSCA, Radiometer) were done simultaneously. All measurements are performed on mechanically ventilated infants. Partial venous carbon dioxide tension (PvCO2) values divided into three groups according to hypocapnia (Group 1: < 4.68 kPa), normocapnia (Group 2: 4.68–7.33 kPa), hypercapnia (Group 3: > 7.33 kPa) and then PvCO2 and PtCO2 data within each group were compared separately.

Results:

A total of 168 measurements of each PvCO2 and PtCO2 data were compared in three separated groups simultaneously (13 in Group 1, 118 in Group 2, and 37 in Group 3). A bias of more than ± 0.7 kPa was considered unacceptable. PtCO2 was related to PvCO2 with acceptable results between the two measurements in hypocapnia (mean difference 0.20 ± 0.19 kPa) and normocapnia (0.002 ± 0.30 kPa) groups. On the other hand in hypercapnia group PtCO2 values were statistically significant (P < 0.001) and lower than PvCO2 data (mean difference 0.81 ± 1.19 kPa)

Conclusions:

PtCO2 measurements have generally good agreement with PvCO2 in hypocapnic and normocapnic intubated infants but there are some limitations especially with high level of CO2 tension. Monitoring of PtCO2 is generally a useful non-invasive indicator of PvCO2 in hypocapnic and normocapnic infants.
Keywords:Blood Gas Monitoring, Transcutaneous   Infant, Newborn   Blood Gas Analysis