SPECT/CT对全髋关节置换术后无菌性假体松动的诊断价值
引用本文:李龙龙,王波,张斌青,郭会利.SPECT/CT对全髋关节置换术后无菌性假体松动的诊断价值[J].医学影像学杂志,2020(2):308-311.
作者姓名:李龙龙  王波  张斌青  郭会利
作者单位:河南中医药大学;河南省洛阳正骨医院(河南省骨科医院)影像中心
基金项目:河南省中医药科学研究专项课题(编号:2018ZY2145)。
摘    要:目的探讨SPECT/CT对全髋关节置换术后无菌性假体松动的诊断价值。方法对临床疑似全髋关节置换术后假体松动的58例患者(69个髋关节),行SPECT/CT检查,观察分析其图像特征。最终结果由手术或5个月以上随访确诊。计算SPECT/CT显像对假体松动的诊断灵敏度、特异度和准确度。结果69个髋关节,临床最终确诊假体松动61个(髋臼松动17个,假体柄松动36个,髋臼及假体柄均松动8个)。SPECT/CT显像阳性64个(其中假阳性3个,余61个与手术结果一致);阴性5个,均为真阴性。SPECT/CT对假体松动的诊断灵敏度、特异度、准确度分别为100%(61/61)、62.50%(5/8)、95.65%(66/69)。结论SPECT/CT显像可灵敏、准确地诊断全髋关节置换术后无菌性假体松动的情况,值得临床推广。

关 键 词:髋关节置换术  假体松动  体层摄影术  发射型计算机  单光子

The diagnostic value of SPECT/CT in aseptic prosthesis looseness after total hip replacement
LI Longlong,WANG Bo,ZHANG Binqing,GUO Huili.The diagnostic value of SPECT/CT in aseptic prosthesis looseness after total hip replacement[J].Journal of Medical Imaging,2020(2):308-311.
Authors:LI Longlong  WANG Bo  ZHANG Binqing  GUO Huili
Institution:(Henan University of Traditional Chinese Medicine,Zhengzhou 450046,P.R.China;Medical Imaging Center,Luoyang Orthopedic Hospital of Henan Province Orthopedic Hospital of Henan Province,Luoyang 471002,P.R.China)
Abstract:Objective To evaluate the diagnostic value of SPECT/CT in aseptic prosthesis looseness after total hip replacement.Methods SPECT/CT was performed in 58 patients(69 hips)who clinically suspected of prosthesis looseness after total hip replacement.We observed and analyzed the image features.Final results were confirmed by surgery or followed up more than 5 months.The diagnostic sensitivity,specificity and accuracy of SPECT/CT for prosthesis looseness were calculated.Results Of the 69 hip joints,61 were finally diagnosed as prosthesis looseness clinically(17 acetabular looseness,36 prosthesis handle looseness,8 acetabular and prosthesis handle looseness).There were 64 positive SPECT/CT images(3 false positive and 61 consistent with the surgical results).5 were negative,all of whom were true negative.The diagnostic sensitivity,specificity and accuracy of SPECT/CT were 100%(61/61),62.50%(5/8)and 95.65%(66/69),respectively.Conclusion SPECT/CT can sensitively and accurately diagnose aseptic prosthesis looseness after total hip arthroplasty.It is worthy of clinical promotion.
Keywords:Hip replacement  Prosthesis looseness  Tomography  emission computed  Single-photon