首页 | 本学科首页   官方微博 | 高级检索  
     

锥形束CT测量牙龈厚度准确性的初步研究
引用本文:贾凯梦,孙晓军,李晶. 锥形束CT测量牙龈厚度准确性的初步研究[J]. 医学影像学杂志, 2020, 0(2): 187-191
作者姓名:贾凯梦  孙晓军  李晶
作者单位:山西医科大学口腔医学院口腔医院;山西医科大学第一医院口腔多学科诊疗中心
基金项目:山西省回国留学人员科研资助项目(编号:2017-112)。
摘    要:目的评价锥形束CT(CBCT)法测量牙龈厚度的可信度及准确性。方法选择20例牙龈健康者的120颗上颌前牙,采用CBCT法和牙龈穿刺法测量唇侧中央龈缘下2mm处牙龈厚度,比较两组差异并用Kappa系数评价两种方法的一致性;应用Kappa系数评价不同检查者用CBCT法测量牙龈厚度的一致性,以及组内相关系数ICC评价同一检查者重复测量数据的可信度。结果CBCT法测量牙龈厚度为(1.23±0.27)mm,牙龈穿刺法测量结果为(1.24±0.26)mm,差异无统计学意义(P>0.05),两组之间一致率为83.33%(Kappa=0.547)。不同的检查者CBCT法测量牙龈厚度的一致性检验Kappa系数为0.814,可信度良好(ICC=0.795),同一检查者CBCT法重复测量数据之间一致性检验Kappa系数为0.864,可信度良好(ICC=0.893);CBCT法对厚牙龈生物型具有高的灵敏度(92.05%)和低的特异度(59.38%)。结论CBCT测量牙龈厚度的结果与牙龈穿刺法的结果较一致,具有良好的可信度。

关 键 词:牙龈厚度  牙龈生物型  上颌前牙  锥形束  体层摄影术,X线计算机

Preliminary study on the accuracy of gingival thickness measurement by cone beam CT
JIA Kaimeng,SUN Xiaojun,LI Jing. Preliminary study on the accuracy of gingival thickness measurement by cone beam CT[J]. Journal of Medical Imaging, 2020, 0(2): 187-191
Authors:JIA Kaimeng  SUN Xiaojun  LI Jing
Affiliation:(Department of Stomatology,Shanxi Medical University School and Hospital of Stomatology,Taiyuan 030001,P.R.China;Oral Multidisciplinary Clinic,The First Hospital of Shanxi Medical University,Taiyuan 030001,P.R.China)
Abstract:Objective To evaluate the reliability and accuracy of gingival thickness measurements by cone-beam computed tomography(CBCT).Methods A total of 120 maxillary anterior teeth from 20 periodontally health volunteers were recruited in this study.The transgingival method and the CBCT were used to measure the gingival thickness at the level of 2 mm apical to the gingival margin midbuccally.The difference was compared between the two methods.Intraoperator and interoperator reliability considering the measurement of CBCT were estimated with Cohen’s kappa coefficient,intraclass correlation coefficient(ICC).Results The agreement rate between CBCT and transgingival method was 83.33%,(Kappa=0.547).There was no statistically significant difference between the two methods.Different measurers used the CBCT method to measure the thickness of the gingival(Kappa=0.814,ICC=0.795)and the same measurer repeated measurement data showed good reliability(Kappa=0.864,ICC=0.893).CBCT method manifested high sensitivity(92.05%)and low specificity(59.38%)for definition of a thick biotype.Conclusion The measurement of gingival thickness with the use of CBCT is consistent with transgingival method and has high reliability.
Keywords:Gingival Thickness  Gingival biotype  Maxillary anterior teeth  Cone-beam  Tomography  X-ray computed
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号