摘    要:为寻求一种简易、快速,成批发测定人血清载脂蛋白B100的方法,作者采用自制羊抗人apoB100血清,以免疫浊度测定法测定了血清apoB100。此法灵敏,特异,易自动化操作,最小检测量为3.1μg,标准曲线测定范围为0.34 ̄1.70g/L,批内及批间重复变异系数分别为1.13% ̄3.48%及2.91% ̄4.48%,测定100例正常成人血清apoB100的含量为0.821±0.165g/L。文内还对

关 键 词:载脂蛋白B100 抗血清 免疫浊度测定法

Quantitation of human apolipoprotein B100 by immunoturbidimetric assay]
B Li,Z Wu,B Liu. Quantitation of human apolipoprotein B100 by immunoturbidimetric assay][J]. Journal of West China University of Medical Sciences, 1997, 28(1): 109-112
Authors:B Li  Z Wu  B Liu
Affiliation:Department of Biochemistry, School of Basic Medical Sciences, Chengdu.
Abstract:An immunoturbidimetric assay(ITA) for quantification of human serum apolipoprotein B100 (apoB100) was developed. The assay is sensitive, rapid, specific, easily automatic and thus convenient for routine work. The minimu measurable concentration of apoB100 was 3.1 micrograms in each assay. The standard curve with a working range of 0.34-1.79 g/L was plotted. The between and within assay coefficients of variation were 1.13%-3.48% and 2.91%-4.48% respectively. The recovery was 99.23% +/- 3.57%. The mean serum concentration of apoB100 in 100 normal subjects was 0.821 +/- 0.165 g/L. The correlation coefficient of RID for apoB100 to ITA was 0.8626(P < 0.001).