首页 | 本学科首页   官方微博 | 高级检索  
     


Thin CaSO4:Dy thermoluminescent dosimeters for calibration of Sr+Y applicators
Authors:Patrí  cia L. AntonioMé  rcia L. Oliveira,Linda V.E. Caldas
Affiliation:a Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, Av. Prof. Lineu Prestes, 2242, 05508-000, São Paulo, SP, Brazil
b Centro Regional de Ciências Nucleares, Comissão Nacional de Energia Nuclear, Av. Prof. Luis Freire, 200, 50740-540, Recife, PE, Brazil
Abstract:Clinical applicators are used in brachytherapy to treat superficial lesions of skin and eye. They should be periodically calibrated according to quality control programs and international recommendations. Thin CaSO4:Dy thermoluminescent dosimeters were used to calibrate various applicators with a dermatological applicator as a reference. The obtained absorbed dose rates were compared with those quoted in their calibration certificates. Depth-dose curves were constructed for all the applicators. A mail dosimetry system was developed for calibration of clinical applicators.
Keywords:90Sr+90Y clinical applicators   Thermoluminescence dosimetry   CaSO4:Dy   Depth-dose curves