体外生物力学评价上胸椎前路钛板内固定装置的三维运动稳定性
作者姓名:罗巨利  詹新立  肖增明  贺聚良  刘会江  赵卫东
作者单位:1广西医科大学第一附属医院脊柱骨病科,广西壮族自治区南宁市 530021 2南方医科大学医学生物力实验室,广东省广州市 510515
基金项目:国家自然科学基金(30960386),课题名称:上胸椎前路内固定系统的研制及相关研究;广西自然科学青年基金(0832039),课题名称:与上胸椎前路内固定相关的解剖及生物力学研究。
摘    要:背景:国内外少见颈前路带锁钢板重建上胸椎的体外生物力学报道。 目的:从生物力学角度评价自行研制的上胸椎前路钛板内固定装置。 方法:15具成人尸体上胸椎标本分成3组,即上胸椎前路钛板内固定组(B),颈前路钛板预弯内固定组(C),颈前路钛板未预弯内固定组(D)。对完整标本进行三维活动度测量,分别安装好各组内固定再次进行标本在前屈、后伸和侧弯及旋转状态下的三维运动稳定性测量,并与完整状态(A)进行对比。 结果与结论:在完整状态下B、C、D 3组标本的三维运动稳定性差异无显著性意义(P > 0.05)。前屈稳定性排列:B>D>C>完整,B、A之间差异有显著性意义(P=0.012);后伸稳定性排列:B>C>A>D,B与A,C,D之间差异有显著性意义(P=0.001,0.021,0.01);旋转状态下的稳定性排列:B>C>A>D,各组差异无显著性意义;侧弯稳定性排列:A>B>C>D,A和C,D之间差异有显著性(P=0.005,0.002)。提示上胸椎前路钛板装置具有较好的三维稳定性,如果采用颈前路钛板内固定建议塑型。

关 键 词:上胸椎  运动范围  内固定  钛板  生物力学  
收稿时间:2010-11-30

Three-dimensional motion stability of upper thoratic anterior titanium plate fixed system: An in vitro biomechanical study
Authors:Luo Ju-li  Zhan Xin-li  Xiao Zeng-ming  He Ju-liang  Liu Hui-jiang  Zhao Wei-dong
Institution:1Department of Orthopedics, the First Affiliated Hospital of Guangxi Medical University, Nanning  530021, Guangxi Zhuang Autonomous Region, China
2Laboratory of Medical Biomechanics, Southern Medical University, Guangzhou  510515, Guangdong Province, China
Abstract:BACKGROUND:There are few reports about in vitro biomechanical studies of anterior cervical locking plate for upper thoracic reconstruction. OBJECTIVE:To evaluate the biomechanical characters of self-designed upper thoratic anterior titanium plate fixed system. METHODS:Intact C7-T6 specimens from 15 fresh corpses were randomly divided into three groups: group B with upper thoracic anterior titanium plate fixation (UTAPF), group C with prespred anterior cervical plate fixation (ACPF), group D with imprespred ACPF. Range of motion (ROM) was tested for the natural intact specimens. Measurement of three-dimensional stability was performed in flexion, extension and lateral bending and rotation state in groups B, C, D after fixation. RESULTS AND CONCLUSION:ROM of the natural intact specimens in the three groups had no significant difference (P > 0.05). The stability of flexion arranged as follows: B>D>C>intact specimens (A). Significant difference between B and A was found (P=0.012). The stability of extension arranged as follows: B>C>A>D, but there was no significant difference among different groups. The stability of rotation arranged B>C>A>D. Between groups no significant difference was found in the rotation condition. The order of stabilities under the lateral bending was A>B>C>D. There were significant differences between A and C as well as between A and D (P=0.005, P=0.002). The findings indicate that the upper thoracic anterior titanium plate has a better biomechanical stability than cervical plate. If cervical anterior plate was applied to upper thoratic spine, plate should be prespred according to kyphosis of the upper thoratic spine.
Keywords: