首页 | 本学科首页   官方微博 | 高级检索  
     


Biomechanical investigation of the primary stability of intramedullary compression nails in the proximal tibia: experimental study using interlocking screws in cryopreserved human tibias
Authors:Hans Gollwitzer  Kiarash Karampour  Matthias Hauschild  Peter Diehl  Raymonde Busch  Wolfram Mittelmeier
Affiliation:(1) Klinik und Poliklinik für Orthopädie und Sportorthopädie der Technischen Universität München, Ismaninger Strasse 22, 81675 München, Germany;(2) Abteilung für Unfallchirurgie, Berufsgenossenschaftliche Unfallklinik Murnau, Prof.-Küntscher-Strasse 8, 82418 Murnau, Germany;(3) Institut für Medizinische Statistik und Epidemiologie der Technischen Universität München, Ismaninger Strasse 22, 81675 München, Germany
Abstract:The biomechanical properties of the proximal tibia with respect to the application of compressible intramedullary nails were examined in a biomechanical model using human cadaver bones. A reference level was examined histomorphometrically in 24 cryopreserved human tibias. Three series of single or double interlocking fixations were tested at various distances from the tibial plateau. Yield strength, rigidity of the bone stock, maximum load, and compression distances were determined during axial strain. The positioning of the interlocking screws had a considerable influence on maximum load and deformation behavior of the bone. The double interlocking tended to display the best stability in the bone stock, and the distally placed single interlocking method was next best. Failure of the bone stock was observed with significant variations in compression distance and load, with great variations between individual specimens. The stability of the bone stock could not be accurately predicted using such clinical parameters as the patientrsquos age, bone dimensions, radiographic parameters, and bone density. When using a tibial compression nail, the proximal interlocking screw should be placed relatively distal to achieve adequate support. It is possible to apply compression with one proximal interlocking screw under low preload or, preferably, with two interlocking screws.
Keywords:Biomechanics  Interlocking nail  Compression  Tibia  Bone density
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏