Contrast timing in computed tomography: effect of different contrast media concentrations on bolus geometry
Authors:Mahnken Andreas H  Jost Gregor  Seidensticker Peter  Kuhl Christiane  Pietsch Hubertus
Affiliation:Department of Diagnostic and Interventional Radiology, RWTH Aachen University, Germany. andreas.mahnken@rwth-aachen.de
Abstract:

Objective

To assess the effect of low-osmolar, monomeric contrast media with different iodine concentrations on bolus shape in aortic CT angiography.

Materials and methods

Repeated sequential computed tomography scanning of the descending aorta of eight beagle dogs (5 male, 12.7 ± 3.1 kg) was performed without table movement with a standardized CT scan protocol. Iopromide 300 (300 mg I/mL), iopromide 370 (370 mg I/mL) and iomeprol 400 (400 mg I/mL) were administered via a foreleg vein with an identical iodine delivery rate of 1.2 g I/s and a total iodine dose of 300 mg I/kg body weight. Time-enhancement curves were computed and analyzed.

Results

Iopromide 300 showed the highest peak enhancement (445.2 ± 89.1 HU), steepest up-slope (104.2 ± 17.5 HU/s) and smallest full width at half maximum (FWHM; 5.8 ± 1.0 s). Peak enhancement, duration of FWHM, enhancement at FWHM and up-slope differed significantly between iopromide 300 and iomeprol 400 (p < 0.05). Except for enhancement at FWHM there were no significant differences between iopromide 300 and iopromide 370 and iopromide 370 and iomeprol 400 (p > 0.05).

Conclusions

Low viscous iopromide 300 results in a better defined bolus with a significantly higher peak enhancement, steeper up-slope and smaller FWHM when compared to iomeprol 400. These characteristics potentially affect contrast timing.
Keywords:Computed tomography   Contrast medium   CT angiography   Contrast enhancement   Iodine concentration