Spectral Entropy and Bispectral Index as Measures of the Electroencephalographic Effects of Sevoflurane
Authors:Ellerkmann  Richard Klaus MD; Liermann  Vidal-Markus&#x;; Alves  Thorsten Michael&#x;; Wenningmann  Ingobert MD&#x;; Kreuer  Sascha MD ; Wilhelm  Wolfram MD&#x;; Roepcke  Heiko MD#; Hoeft  Andreas MD  PhD; Bruhn  Jrgen MD#
Institution:Ellerkmann, Richard Klaus M.D.*; Liermann, Vidal-Markus?; Alves, Thorsten Michael?; Wenningmann, Ingobert M.D.?; Kreuer, Sascha M.D.§; Wilhelm, Wolfram M.D.∥; Roepcke, Heiko M.D.#; Hoeft, Andreas M.D., Ph.D.**; Bruhn, Jörgen M.D.#
Abstract:Background: Recently, entropy algorithms have been proposed as electroencephalographic measures of anesthetic drug effects. Datex-Ohmeda (Helsinki, Finland) introduced the Entropy Module, a new electroencephalographic monitor designed for measuring depth of anesthesia. The monitor calculates a state entropy (SE) computed over the frequency range of 0.8-32 Hz and a response entropy (RE) computed over the frequency range of 0.8-47 Hz. The authors investigated the dose-response relation of SE and RE during sevoflurane anesthesia in comparison with the Bispectral Index (BIS).

Methods: Sixteen patients were studied without surgical stimulus. Anesthesia was induced by sevoflurane inhalation with a tight-fitting facemask. Sevoflurane concentrations were increased and subsequently decreased and increased two to four times until the measurement was stopped and patients were intubated for surgery. The performances of SE, RE, and BIS to predict the estimated sevoflurane effect site concentration, obtained by simultaneous pharmacokinetic and pharmacodynamic modeling, were compared by calculating the correlation coefficients and the prediction probability.

Results: State entropy, RE, and BIS values decreased continuously over the observed concentration range of sevoflurane. Correlation coefficients were slightly but not significantly better for entropy parameters (0.87 +/- 0.09 and 0.86 +/- 0.10 for SE and RE, respectively) than for BIS (0.85 +/- 0.12). Calculating the prediction probability confirmed these results with a prediction probability of 0.84 +/- 0.05 and 0.82 +/- 0.06 for SE and RE, respectively, and 0.80 +/- 0.06 for BIS.